Intraoral scanner tips are small components, but they play a big role in image quality, patient comfort, and infection control. As digital dentistry adoption continues to grow, dental clinics and labs are paying more attention to how scanner tips should be cleaned, sterilized, and reused safely—especially as more manufacturers introduce reusable scanning tips to reduce ongoing costs.
Whether you are using a system with disposable tips or a reusable model such as UP3D's Clariscan UP610, maintaining proper sterilization protocols is essential for both clinical safety and long-term scanner performance.
Below is a practical, clinician-friendly guide to help you clean, disinfect, and reuse intraoral scanner tips correctly.

Why Proper Cleaning and Sterilization Matter
A scanner tip enters the patient's mouth, which means it must be treated with the same level of infection control as any other intraoral device. Proper care affects:
-
Patient safety – prevents cross-contamination
-
Image quality – residue can cause scanning artifacts or fogging
-
Longevity of the tip – incorrect cleaning can warp optics or damage coatings
-
Workflow efficiency – having clean tips ready avoids appointment delays
-
Cost control – reusable tips save money only if maintained properly
Reusable tips are becoming increasingly popular because they offer cost advantages and reduce disposable waste. But they only remain safe if sterilized according to guidelines.
Step 1: Pre-Cleaning Immediately After Use
Right after removing the scanner from the patient's mouth:
-
Wipe the tip with a soft, lint-free cloth
-
Remove saliva, debris, and blood residue
-
Avoid abrasive materials that could scratch the optical window
-
Detach the tip (if applicable) and prepare it for disinfection
Early pre-cleaning prevents buildup that becomes harder to remove later.
Step 2: Disinfection Before Sterilization
Most manufacturers recommend an intermediate disinfection step.
-
Use approved medical surface disinfectants (non-corrosive, alcohol-compatible).
-
Avoid strong solvents like acetone or chlorinated solutions.
-
Ensure disinfectant contact time follows the label instructions.
-
Dry the tip completely before sterilizing.
This step reduces bioburden and protects the sterilizer from contamination.
Step 3: Sterilization — Methods That Are Safe for Scanner Tips
This is where many clinics make mistakes. Not all scanner tips tolerate heat, and not all models can be autoclaved. Always check your scanner's IFU (Instructions for Use).
Below are the most common and safe methods:
1. Steam Autoclave (Most Common)
Many reusable scanner tips, including the Clariscan UP610 tip, are autoclavable at 121–134°C depending on the cycle.
Best practices:
-
Place tips in sterilization pouches
-
Do not overload the chamber
-
Use validated cycles only
-
Allow the tip to cool before handling or installing
2. Low-Temperature Plasma Sterilization
Ideal for heat-sensitive optical components.
Advantages:
-
No heat damage
-
No moisture
-
Short cycles
Check compatibility first—some tips are approved for plasma sterilization only.
3. Dry Heat Sterilization (If Approved)
A minority of tips allow dry heat sterilization.
Always avoid:
-
Ultra-high temperatures
-
Rapid reheating/cooling
-
Any method not listed in the IFU
Step 4: Proper Storage After Sterilization
To maintain sterility:
-
Store tips in sealed pouches
-
Keep them in a clean, dry cabinet
-
Avoid touching optical surfaces
-
Rotate tips (if you have multiple) to reduce wear
This helps maintain lifespan and prevents accidental contamination.
Step 5: How Many Times Can a Reusable Tip Be Sterilized?
Reusable scanner tips typically withstand 100–150 sterilization cycles, though this varies by manufacturer.
Signs a tip should be replaced:
-
Scratches on the optical window
-
Fogging or hazy images
-
Warping from heat
-
Cracks or discoloration
-
Distorted scan data
A tracking system (simple spreadsheet or QR code labeling) helps monitor sterilization cycles.
Common Cleaning Mistakes to Avoid
-
Using abrasive brushes or metal tools
-
Soaking the tip in strong chemicals
-
Using high-pressure steam beyond recommended limits
-
Skipping drying steps
-
Attempting “quick clean” shortcuts between patients
-
Reassembling while still warm (can warp plastics)
These mistakes can dramatically shorten tip lifespan and compromise imaging accuracy.
Cost and Sustainability Benefits of Reusable Tips
Reusable tips are becoming standard because they help clinics:
-
Reduce monthly consumable spending
-
Lower medical waste
-
Improve environmental impact
-
Maintain consistent scan performance
-
Support high-volume clinical workflows
Clinics using scanners like the UP610 often report significant savings within the first year.
Final Thoughts
Cleaning, sterilizing, and reusing intraoral scanner tips safely is a key part of building a reliable digital workflow. When done correctly, it protects patients, preserves your equipment, and reduces long-term operating costs.
Whether your clinic uses disposable or reusable tips, following evidence-based hygiene protocols ensures that every scan is safe, clear, and clinically reliable.